Healthcare Asia Pacific
Home > Events & Announcements > Omron NEU-200 - Information for accompanying documents in the scope of IEC 60601-1-2:2014

Omron NEU-200 - Information for accompanying documents in the scope of IEC 60601-1-2:2014

By Omron Healthcare India Pvt. Ltd., 16 Apr 2021

OMRON mains-operated Nebulizer NE-U200

Applicable models: Refer to” EMC product reference table Nebulizer” Information for accompanying documents in the scope of IEC 60601-1-2:2014

Important information regarding Electro Magnetic Compatibility (EMC):


This device manufactured by OMRON HEALTHCARE Co., Ltd. conforms to IEC 60601-1-2:2014 Electro Magnetic Compatibility (EMC) standard. Nevertheless, special precautions need to be observed:

The use of accessories other than those specified or provided by OMRON could result in increased electromagnetic emission or decreased electromagnetic immunity of the device and result in improper operation.
The use of the device adjacent to or stacked with other device should be avoided because it could result in improper operation. In case such use is necessary, the device and other device should be observed to verify that they are operating normally.
Portable RF communications device (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm  to any part of the device, specified by OMRON. Otherwise, degradation of the performance of the device could result.
Refer to further guidance below regarding the EMC environment in which the device should be used.


Table 1 - EMISSION Limits and Compliance

Phenomenon

EMISSION Limits

Compliance

Radiated RF EMISSIONS

CISPR 11

Group1, Class B

Voltage fluctuations and flicker

See IEC 61000-3-3

       N/A

 

Table 2 ‐ IMMUNITY TEST LEVELS

Phenomenon

Basic EMC standard

IMMUNITY TEST LEVELS

Electrostatic discharge

IEC 61000‐4‐2

±8 kV contact

±2 kV,±4 kV,±8 kV,±15 kV air for enclosure port

Radiated RF electromagnetic fields

IEC 61000‐4‐3

10 V/m

80 MHz to 2.7 GHz

80 % AM at 1 kHz for enclosure port

Proximity fields from RF wireless communications equipment

IEC 61000‐4‐3

 

 

See table 3

Electrical fast transients / bursts

IEC 61000‐4‐4

                             N/A

Surges Line‐to‐line

IEC 61000‐4‐5

                             N/A

Conducted disturbances induced by RF fields

IEC 61000‐4‐6

                             N/A

Rated power frequency magnetic fields

IEC 61000‐4‐8

30 A/m

50 Hz or 60 Hz for enclosure port

Voltage dips

IEC 61000‐4‐11

                 N/A

Voltage interruptions

IEC 61000‐4‐11

                N/A

Note: UT is the A.C. mains voltage prior to application of the test level.

 

Table 3 - Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications device

Test frequency (MHz)

Band (MHz)

Service

Modulation

Maximum power (W)

Distance (m)

IMMUNITY TEST LEVEL

(V/m)

385

380 to 390

TETRA 400

Pulse modulation 18 Hz

1.8

0.3

27

 

450

 

430 to 470

 

GMRS 460,

FRS 460

FM

5 kHz deviation 1 kHz sine

 

2

 

0.3

 

28

710

 

704 to 787

 

LTE Band 13, 17

 

Pulse modulation 217 Hz

 

0.2

 

0.3

 

9

745

780

810

 

 

800 to 960

GSM 800/900, TETRA 800,

iDEN 820,

CDMA 850,

LTE Band 5

 

 

Pulse modulation 18 Hz

 

 

2

 

 

0.3

 

 

28

870

930

1720

 

 

 

1700 to 1990

GSM 1800;

CDMA 1900;

GSM 1900; DECT;

LTE Band 1, 3,

4, 25; UMTS

 

 

Pulse modulation 217 Hz

 

 

 

2

 

 

 

0.3

 

 

 

28

1845

1970

 

 

2450

 

 

2400 to 2570

Bluetooth, WLAN,

802.11 b/g/n , RFID 2450, LTE Band 7

 

 

Pulse modulation 217 Hz

 

 

2

 

 

0.3

 

 

28

5240

 

5100 to 5800

 

WLAN 802.11

a/n

 

Pulse modulation 217 Hz

 

0.2

 

0.3

 

9

5500

5785

 

Return to Events Page